Milwaukee

Milwaukee Waterproof EC Meter (C66) Manual Calibration

PHEC0132

Sold by Natural Systems. More information

The C66 EC Meter is a pocket-sized, waterproof Electro Conductivity (EC) meter.

The C66 EC Meter measures with temperature compensation and has a replaceable probe (MA73076).

The C66 EC Meter has an ON button (ON) and an OFF button (OFF).< /p>

The C66 EC Meter has a very wide measurement scale, measuring EC from 0.00 mS to 10.00 mS on its LCD screen.

Operation:

  • Remove the protection; Turn on the C66 meter by pressing the ON button; Immerse the meter to the limit in the liquid to measure. Wait a few seconds for the data to stabilize.
  • After use, wash the polo shirts with water and dry them.
  • Always replace the cap after use.
  • C66 measures from 0.00 to 10.00 mS/cm. If the reading goes over 10.00, the meter will read "1".

Calibration:

  • Unscrew the battery cap (top of the meter) to access the calibration screw.
  • Immerse the probe to the limit in EC 1413 calibration liquid. Wait for the data to stabilize and adjust to 1.41 with the screwdriver.
  • Clean the probe with tap water and let the poles dry.

Change electrode:

Remove the meter cover. Unscrew the black ring and take it out. Remove the probe and replace it with a new one, with Be careful not to damage the connections.

Changing batteries:

The C66 gives a low battery warning by turning itself off. The batteries are located at the top of the meter. unscrew and switch to 3 x 1.5V.